实验方案(Protocols)

Factor VII Related Antigen Content Assay

Summary

The turbidity of the immunocomplexes between coagulation factor VIII-related antigen (ⅧR; Ag) and its corresponding specific antibody is determined by the transmission method, and the turbidity is proportional to the concentration of ⅧR:Ag in the serum. This experiment is from the laboratory instruction of Mudanjiang Medical College undergraduate 5-year testing program.

Operation method

Factor VII-related antigen content assay

Principle

The turbidity of the immunocomplexes between coagulation factor VIII-related antigen (VIIIR; Ag) and its corresponding specific antibody is determined by transmission assay and is proportional to the serum concentration of IV VIIIR:Ag.

Move

Experimental method: manual, semi-automatic measurement


Mixing 37 ℃ water bath for 20 minutes, 340 nm with antiserum zero point determination of each tube A value D

Data processing: ⅧR: Ag content = (A-value of measurement tube / A-value of reference tube)×100%.

Normal value: 60-150%.

Caveat

1. When the turbidity of the specimen is too high, the dilution ratio of the specimen should be regulated and re-specified.

2. high blood fat. Jaundice and mild hemolysis have little effect on the results of this method.

3. Linear range of the method: 0-200%.Clinical significance: ⅧR: Ag content increased: seen in coronary heart disease myocardial infarction, cerebral thrombosis kidney disease, etc.. Decrease in ⅧR:Ag level: seen in vascular pseudohaemophilia, haemophilia carriers, etc.


For more product details, please visit Aladdin Scientific website.

https://www.aladdinsci.com/

目录: 实验方案(Protocols)
探索主题: Clinical trial

Da — 若无特别说明,分子量单位默认为道尔顿。   Mw — 重均分子量。   Mn — 数均分子量。

产品仅供科研与开发使用,不用于人体、动物、诊断或治疗。

引用本文

阿拉丁科学.《Factor VII Related Antigen Content Assay》. 阿拉丁知识库,更新于 2026年7月25日。 https://www.aladdin-e.com/zh_cn/faqs/factor-vii-related-antigen-content-assay-en.html
这篇文章对您有帮助吗? Yes No 有 5 人觉得有帮助