LC–MS/MS separation and quantitation of ribavirin in chicken and comparison of different mass spectrometric platforms

Journal 2023 · IF 4.6 · 17(1): 1-10 · 发表于 2023 年 8 月 7 日
作者: Xu Daokun,Huang Haolun,Hu Wenyan,Liu Xinmei,Yang Jun
阅读全文 在 PubMed 查看

摘要

A liquid chromatography–tandem mass spectrometry (LC–MS/MS) method was developed for the analysis of ribavirin in chicken. Samples was extracted with 0.1% formic acid and purified by Hypercarb cartridge prior to LC–MS/MS analysis. The eluates were evaporated to dryness, reconstituted in 1 mL 5mM ammonium acetate containing 5% acetonitrile (v/v) and 0.1% (v/v) formic acid. Chromatographic separation was performed on a Hypercarb analytical column under a gradient elution program with acetonitrile and 0.1% (v/v) formic acid in 5 mM ammonium acetate at a flow rate of 0.6 ml/min. The intraday and interday accuracy ranged from − 7.83 − 1.39%, and − 6.38 − 2.25%, with precisions between 1.34 − 3.88%% and 1.10 − 4.67%. The limits of detection (LODs) and limits of quantitation (LOQs) of ribavirin was 0.1 ng/mL and 0.5 ng/mL, respectively. The method was validated for linearity, accuracy, precision, matrix effect and stability. Application of the method confirmed 3 ribavirin positive samples out of 50 commercial chicken samples, with concentrations of ribavirin ranging from 0.9 μg/kg to 5.8 μg/kg a, respectively. Additionally, both AB Sciex 5500 and Agilent 6945B were proven to be suitable in ribavirin separation and quantification. The described method is suitable for the determination of ribavirin in chicken in analytical practice to monitor illegal addition of this kind of anti–viral drug.

本研究中使用的阿拉丁产品

13 款产品在本文中被引用