已核实 DOI 相关产品: 28

Chiral organic-silica hybrid monolithic HPLC column with dynamic grafting BSA for the enantiomeric separation

Journal 2023 · IF 4.8 · 发表于 2023 年 7 月 19 日
作者: Yue-Na Sun,Lei Shi,Mei-Li Shi,Li-Qing Yu,Hong-Yuan Yan,Yun-Kai Lv
阅读全文 在 PubMed 查看

摘要

BSA functionalized hybrid monolithic column prepared by in-situ dynamic grafting on traditional stainless steel column for enantiomeric separation. A hybrid silica monolithic column (HMC) was prepared using a single precursor (methytrimethoxysilane) in stainless steel HPLC column (50 mm × 4.6 mm) by sol–gel method. The chiral hybrid monolithic column (CHMC) was prepared by grafting 3-aminopropyl trimethoxysilane onto HMC, and then reacting bovine serum albumin (BSA) with Glutaraldehyde. The preparation conditions were optimized, and the maximum grafting amount of BSA was 48.6 mg g −1 . The CHMC was characterized and its chromatographic performance was tested. The results showed that the CHMC has a homogeneous network skeleton structure, good stability and reproducibility. The CHMC was combined with HPLC, under optimized chromatographic conditions, the baseline separation of tryptophan and pidotimod enantiomers were successfully achieved.

本研究中使用的阿拉丁产品

28 款产品在本文中被引用