已核实 DOI PubMed 相关产品: 166

Fiber SPR biosensor sensitized by MOFs for MUC1 protein detection

Journal 2023 · IF 6.1 · 发表于 2023 年 3 月 20 日
作者: Rui Wang,Chunlan Liu,Yong Wei,Ze Ran,Tianci Jiang,Chunbiao Liu,Chen Shi,Zhuo Ren,Xingkai Wang,Zhihai Liu,Yu Zhang
阅读全文 在 PubMed 查看

摘要

The concentration of tumor markers is low, which needs a highly sensitive, stable and fast detection method. In this paper, we proposed and demonstrated a U-shape fiber SPR biosensor sensitized by MOFs materials. The surface of the U-shape SPR sensor was modified with MOFs materials to enhance the sensitivity, and the nucleic acid aptamer was immobilized on the sensor surface because of the biocompatibility of MOFs materials. By the high specificity of the nucleic acid aptamer, the MUC1 protein was recognized and detected. The testing results indicate that the sensor has a logarithmic linear response in the MUC1 protein concentration detection range of 1 pg/ml-100 μg/ml, its sensitivity and detection limit are 5.33 nm/log(μg/ml) and 0.16 pg/ml respectively. After being sensitized by MOFs, the detection sensitivity of the sensor can be increased by 1.62 times,the LOD can be decreased by 0.75 times. The sensor has high sensitivity and specificity, which has broad application prospects in clinical detection of tumor markers.

本研究中使用的阿拉丁产品

166 款产品在本文中被引用