已核实 DOI PubMed 相关产品: 166

A wearable electrochemical sensor based on β-CD functionalized graphene for pH and potassium ion analysis in sweat

Journal 2022 · IF 6.556 · 发表于 2022 年 4 月 16 日
作者: Xiaoqing Cui,Yu Bao,Tingting Han,Zhenbang Liu,Yingming Ma,Zhonghui Sun
阅读全文 在 PubMed 查看

摘要

Here, we developed a wearable electrochemical sensor for pH and K + monitoring in sweat. The sensor was composed of flexible reference electrode, pH response electrode and K + selective electrode, which were prepared through printing β-CD functionalized graphene (β-CD/RGO) water suspension on conductive PET substrate with microelectronic printer. β-CD/RGO not only served as the pH sensitive material for pH response electrode with good sensitivity, selectivity and reproducibility due to its abundant oxygen-containing functional groups, but also worked as the ion-to-electron transducer for K + selective electrode with good sensitivity. The wearable sensor exhibited good potential stability at different bending states. On-body sweat pH and K + measurements showed high accuracy compared with ex-situ analysis.

本研究中使用的阿拉丁产品

166 款产品在本文中被引用