Residual dipolar couplings (RDCs) are widely used in NMR for the investigation of protein structure and dynamics. However, commonly used RDC media ( e.g. , C 12 E 5 /Hexanol/H 2 O or PEG system, lipid bilayer or bicelle system) are unstable viscous liquids, making it challenging to obtain stable and homogeneous NMR samples by laborious manual operation. In this paper, we report a unique fishhook-shaped micromixer (FS-mixer) that enables fast and thorough mixing of high-viscosity PEG and bicelle media for RDC analysis. Numerical simulations and experimental characterizations proved that the micromixer could achieve a complete mixing of viscous liquids in 382.4 μs, improved significantly over current mixers. In compared to manually prepared RDC samples, the on-chip prepared samples have a better uniformity and alignment efficiency, and show more stable peak split in the 2 H spectrum. As a proof of concept, we prepared a RDC sample for INSM1 ZF1-2 protein by using this FS-mixer. The IPAP-HSQC spectra demonstrated that this strategy can efficiently prevent the generation of low resolution and distorted spectra, which could facilitate further refinement of the NMR structure. This micromixer-assisted sampling method greatly simplifies the procedure, increases the media uniformity and stability, and shortens the time in RDC analysis. We believe it could serve as a useful tool in RDC-based NMR analysis.