A mathematical method for calibrating the signal drift in liquid chromatography - mass spectrometry analysis

Journal 2021 · IF 6.057 · 发表于 2021 年 5 月 13 日
作者: Yu-Ning Hu,Jin-Tao Zhan,Quan-Fei Zhu,Ting Hu,Na An,Zhen Zhou,Yong Liang,Wei Wang,Zhi Han,Jie Wang,Fu-Qiang Xu,Yu-Qi Feng
阅读全文 在 PubMed 查看

摘要

Liquid chromatography coupled with high-resolution mass spectrometry (LC-HRMS) has become the most versatile analytical tool for profiling small-molecule compounds and increasingly been applied in many fields. Nevertheless, LC-MS based quantification still face some challenges, such as signal drift in LC-MS, which may affect the validity of the obtained data and lead to misinterpretation of biological results. Here, we established a calibration method known as “RIM” to compensate the signal drift of LC-MS. To this end, a mixture of d 4 -2-dimethylaminoethylamine ( d 4 -DMED)-coded normal fatty acids (C5–C23) was used as calibrants to construct RIM calibration. With the addition of calibrants, not only the MS signal drift, but also the mass accuracy and LC retention time can be calibrated, thereby improving the reliability of quantitative data. The effectiveness of RIM was carefully validated using a human serum extract spiked with 34 standards and then RIM was applied for rat brain untargeted metabolome research. In addition, to expand the functionality and flexibility of RIM for data handling, we generated a MATLAB-based RIM program, which implements the above concepts and allows automatic data process. Taken together, the proposed RIM method has potential application in large-scale quantitative study of complex samples.