已核实 DOI PubMed 相关产品: 63

A flexible rGO electrode: a new platform for the direct voltammetric detection of salicylic acid

Journal 2020 · IF 2.596 · 12(31): 3892-3900 · 发表于 2020 年 6 月 5 日
作者: Qinghua Yu,Ying Zhao,Lihui Huang,Jiawei Sun,Dangqin Jin,Yun Shu,Qin Xu,Xiao-Ya Hu
阅读全文 在 PubMed 查看

摘要

Flexible sensors are of considerable interest for the development of wearable smart miniature devices. This work reported a flexible electrochemical platform based on reduced graphene oxide (rGO) for the detection of salicylic acid (SA). The free-standing and flexible rGO electrode was prepared via a simple extruded process. Dynamic mechanical deformation and bending studies illustrated the resilience and compliance of the flexible electrode against extreme mechanical deformations. Quantitative analysis of SA was performed by using differential pulse voltammetry (DPV) with this flexible rGO electrode. Linearity ranges for SA were obtained from 1.0 × 10−10 M to 1.0 × 10−5 M with the detection limit of 2.3 × 10−11 M (S/N = 3). This strategy provided a new insight into the design and application of flexible electrodes. It will extend the applications of rGO in sensing, bio-electronics and lab-on-chip devices.

本研究中使用的阿拉丁产品

63 款产品在本文中被引用