已核实 DOI PubMed 相关产品: 10

A time-resolved luminescence aptasensor of ofloxacin based on rolling circle amplification and magnetic separation

Journal 2020 · IF 3.637 · 412(19): 4555-4563 · 发表于 2020 年 5 月 29 日
作者: Huang Yukun,Wang Chong,Huo Qiuyu,Song Yaning,Du Gang,Wang Lijun,Yang Xiao,Chen Xianggui
阅读全文 在 PubMed 查看

摘要

A novel sensitive, competitive, and time-resolved luminescence sensor for the detection of ofloxacin (OFL) was developed in this study. The sensor used OFL-specific aptamer as a recognition molecule and rolling circle amplification (RCA) as a signal amplification tool. In this way, the time-resolved luminescence can not only avoid background noise from sample, but also provide robust luminescence for detection. Besides, the separation and enrichment of target veterinary drug can be conducted assisted by magnetic treatment. Under optimal conditions, the logarithmic correlation between the concentration of OFL and the luminescence intensity was found to be linear in the range of 5 × 10 −11 –5 × 10 −8  mol L −1 ( R 2  = 0.9988), with a detection limit (LOD) of 32.1 pmol L −1 . Furthermore, this method was applied to the determination of OFL in chicken and pork samples, exhibiting good recovery (72.5–100%) and repeatability (RSD < 10.0%). These results confirm that this novel established method has good application potential for the detection of OFL in food samples.

本研究中使用的阿拉丁产品

10 款产品在本文中被引用