Electrochemical sensor for Cd2+ and Pb2+ detection based on nano-porous pseudo carbon paste electrode

Journal 2019 · IF 3.839 · 发表于 2019 年 5 月 8 日
作者: Yuan Liu,Taotao Li,Chuxuan Ling,Zhu Chen,Yan Deng,Nongyue He
阅读全文 在 PubMed 查看

摘要

An electrochemical sensor based on self-made nano-porous pseudo carbon paste electrode (nano-PPCPE) has been successfully developed, and used to detect Cd 2+ and Pb 2+ . The experimental results showed that the electrochemical performance of nanoPPCPE is evidently better than both glassy carbon electrode (GCE) and pure carbon paste electrode (CPE). Then the prepared nano-PPCPE was applied to detect Cd 2+ and Pb 2+ in standard solution, the results showed that the electrodes can quantitatively detect trace Cd 2+ and Pb 2+ , which has great significance in electrochemical analysis and detection. The linear ranges between the target ions concentration and the DPASV current were from 0.1–3.0 μmol/L, 0.05–4.0 μmol/L for Cd 2+ and Pb 2+ , respectively. And the detection limits were 0.0780 μmol/L and 0.0292 μmol/L, respectively. Moreover, the preparation of the nano-PPCPE is cheap, simple and has important practical value.

本研究中使用的阿拉丁产品

17 款产品在本文中被引用