已核实 DOI PubMed 相关产品: 5

Turbidimetric inhibition immunoassay revisited to enhance its sensitivity via an optofluidic laser

Journal 2019 · IF 9.518 · 发表于 2019 年 2 月 19 日
作者: Xi Yang,Wenxiong Shu,Yanqiong Wang,Yuan Gong,Chaoyang Gong,Qiushu Chen,Xiaotian Tan,Gang-Ding Peng,Xudong Fan,Yun-Jiang Rao
阅读全文 在 PubMed 查看

摘要

Turbidimetric inhibition immunoassay (TIIA) is a classic immunodiagnostic method that has been extensively used for biomarker detection. However, the low sensitivity of this technique hinders its applications in the early diagnosis of diseases. Here, a new concept, optofluidic laser TIIA (OFL-TIIA), is proposed and demonstrated for sensitive protein detection. In contrast to the immunoreaction in traditional TIIA, in which the single-pass laser loss is detected, the immunoreaction in the OFL-TIIA method takes place in a laser cavity , which considerably increases the loss induced by antigen-antibody complexes (AACs) via the amplification effect of the laser. A commercial IgG TIIA kit was selected as a demonstrative model to characterize the performance of OFL-TIIA. A wide dynamic range of five orders of magnitude with an exceptional limit of detection (LOD) (1.8 × 10 −10 g/L) was achieved. OFL-TIIA is a fast, sensitive, and low-cost immunoassay with a simple homogeneous and wash-free process and low-volume sample consumption, thus providing a new detection platform for disease diagnostics.

本研究中使用的阿拉丁产品

5 款产品在本文中被引用