A molecularly imprinted polymer based chemiluminescence array sensor for one-step determination of phenothiazines and benzodiazepines in pig urine

Journal 2018 · IF 1.671 · 34(1): 98-105 · 发表于 2018 年 12 月 12 日
作者: Wan Qiu Xia,Hui Cai Zhang,Geng Nan Wang,Jing Liu,Jian Ping Wang
阅读全文 在 PubMed 查看

摘要

The residues of phenothiazines and benzodiazepines in foods of animal origin are dangerous to consumers. For inspection of their abuses, this study for the first time reported on the use of a chemiluminescence array sensor for the simultaneous determination of four phenothiazines and five benzodiazepines in pig urine. Two molecularly imprinted polymers were coated in different wells of a conventional 96-well microtiter plate as the recognition reagents. After sample loading, the absorbed analytes were initiated directly by using an imidazole enhanced bis(2,4,6-trichlorophenyl)oxalate–hydrogen peroxide system to emit light. The assay process consisted of only one sample-loading step prior to data acquisition, so one test was finished within 10 min. The limits of detection for the nine drugs in the pig urine were in a range of 0.1 to 0.6 pg/mL, and the recoveries from the fortified blank urine samples were in a range of 80.3 to 95%. Furthermore, the sensor could be reused six times. Therefore, this sensor could be used as a simple, rapid, sensitive and reusable tool for routine screening for residues of phenothiazines and benzodiazepines in pig urine.

本研究中使用的阿拉丁产品

3 款产品在本文中被引用