已核实 DOI PubMed 相关产品: 2

Determination of nitenpyram and 6-chloronicotinic acid in environmental samples by ion chromatography coupled with online photochemically induced fluorescence detector

Journal 2018 · IF 2.415 · 41(22): 4096-4104 · 发表于 2018 年 9 月 19 日
作者: Nadeem Muhammad,Yun Zhang,Weixia Li,Yong-Gang Zhao,Amjad Ali,Qamar Subhani,Tariq Mahmud,Junwei Liu,Hairong Cui,Yan Zhu
阅读全文 在 PubMed 查看

摘要

A simple, cost-effective, sensitive, and quick method for the determination of nitenpyram and its metabolite 6-chloronicotinic acid in environmental samples was developed by coupling an ion chromatograph with a fluorescence detector and a post-column photochemical reactor. This developed analytical method involved a rapid sample extraction by modified and miniaturized quick, easy, cheap, effective, rugged, and safe method followed by isocratic ion chromatographic separation of nitenpyram and 6-chloronicotinic acid into an IonPac ™ AS11-HC column protected by IonPac ™ AG11A guard column by running 30 mM NaOH + 10% acetonitrile mobile phase. A homemade post-column photochemical reactor was also integrated with the ion chromatographic system for online transformation of both analytes into their respective highly fluorescent photoproduct in basic media without using an extra pump. The developed method was validated by following SANTE/11945/2015 guidelines on analytical quality control and validation procedures. The method showed a good linear response (r > 0.999), improved limit of detection (0.101–0.132 μg/L), minimum or no matrix effect, excellent recoveries (90.2–100.10%) and relative standard deviations were found to be ≤6.50%.

本研究中使用的阿拉丁产品

2 款产品在本文中被引用