已核实 DOI PubMed 相关产品: 15

A non-enzymatic electrochemical immunoassay for quantitative detection of Escherichia coli O157:H7 using Au@Pt and graphene

Journal 2018 · IF 2.275 · 发表于 2018 年 8 月 23 日
作者: Fanjun Zhu,Guangying Zhao,Wenchao Dou
阅读全文 在 PubMed 查看

摘要

Herein, a non-enzymatic sandwich-type electrochemical immunoassay was fabricated for quantitative monitoring of Escherichia coli O157:H7 ( E. coli O157:H7). Silica coated Fe 3 O 4 magnetic nanoparticles (Fe 3 O 4 @SiO 2 ) were modified with mouse anti- E. coli O157:H7 monoclonal antibody (Ab 1 ) to act as capture probes to reduce detection time and increase the sensitivity of the immunoassay . The Au@Pt nanoparticles were loaded on neutral red (NR) functionalized graphene to form composite complex rGO-NR-Au@Pt. rGO-NR-Au@Pt has high specific surface area and good biocompatibility. rGO-NR-Au@Pt was used as the carriers of detection antibodies (Ab 2 ). Au@Pt catalyzed the reduction of hydrogen peroxide (H 2 O 2 ) to detection of E. coli O157:H7 with the thionine (TH) as electron mediator to effectually amply the current signal. Under the optimized conditions, a linear relationship between the reduction peak current change (ΔI pc ) and the logarithm of the E. coli O157:H7 concentration is obtained in the range from 4.0 × 10 3 to 4.0 × 10 8   CFU  mL −1 and the limit of detection (LOD) is 4.5 × 10 2  CFU mL -1  at a signal-to-noise ratio of 3. The immunoassay exhibits acceptable specificity, reproducibility and stability on the detection of E. coli O157:H7. Furthermore, the immunoassay showed good performance in pork and milk samples. The results suggest that this immunoassay will be promising in the food safety area.

本研究中使用的阿拉丁产品

15 款产品在本文中被引用