A ratiometric electrochemiluminescent biosensor for Con A detecting based on competition of dissolved oxygen

Journal 2018 · IF 8.173 · 发表于 2018 年 8 月 6 日
作者: Han Zhang,Cong Zhang,Di Liu,Fumei Zuo,Shihong Chen,Ruo Yuan,Wenju Xu
阅读全文 在 PubMed 查看

摘要

A novel ratiometric electrochemiluminescent (ECL) biosensor was designed for the detection of concanavalin A (Con A) based on two ECL emitters competing for the dissolved oxygen (O 2 ). In this strategy, CdTe quantum dots (QDs) were used as the cathodic emitter and N-(aminobutyl)-N-(ethylisoluminol) (ABEI) was used as the anodic emitter. With the presence of dissolved O 2 utilized as co-reactant, CdTe QDs showed an ECL emission at − 1.7 V, and ABEI showed an emission at + 0.6 V. Phenoxy-derivatized dextran (Dexp), as a recognition element, was immobilized by graphene oxide functionalized CdTe QDs (G-CdTe QDs) to recognize Con A, and further combine with Dexp, gold and platinum nanoparticles decorated ABEI (Dexp-Au-Pt-ABEI) to form a sandwich structure. With the increasing concentration of analyte Con A, the ECL signal of cathodic CdTe QDs decreased and the anodic response of ABEI increased, thus achieving a ratiometry detection of Con A with a wide linear range from 1.0 × 10 −4 ng/mL to 10 ng/mL. The detection limit was low to 3.0 × 10 −5 ng/mL. This proposed ratiometric ECL biosensor not only conquered the false errors caused by external interferences, but excluded the disability caused by exogenous co-reactant via the application of common dissolved O 2 . It also exhibited an excellent stability, selectivity and reproducibility.

本研究中使用的阿拉丁产品

101 款产品在本文中被引用