Sensitive paper-based analytical device for fast colorimetric detection of nitrite with smartphone

Journal 2018 · IF 3.307 · 410(11): 2665-2669 · 发表于 2018 年 3 月 19 日
作者: Zhang Xiu-Xiu,Song Yi-Zhen,Fang Fang,Wu Zhi-Yong
阅读全文 在 PubMed 查看

摘要

On-site rapid monitoring of nitrite as an assessment indicator of the environment, food, and physiological systems has drawn extensive attention. Here, electrokinetic stacking (ES) was combined with colorimetric reaction on a paper-based device (PAD) to achieve colorless nitrite detection with smartphone. In this paper, nitrite was stacked on the paper fluidic channel as a narrow band by electrokinetic stacking. Then, Griess reagent was introduced to visualize the stacking band. Under optimal conditions, the sensitivity of nitrite was 160-fold increased within 5 min. A linear response in the range of 0.075 to 1.0 μg mL −1 ( R 2  = 0.99) and a limit of detection (LOD) of 73 ng mL −1 (0.86 μM) were obtained. The LOD was 10 times lower than the reported PAD, and close to that achieved by a desktop spectrophotometer. The applicability was demonstrated by nitrite detection from saliva and water with good selectivity, adding 100 times more concentrated co-ions. High recovery (91.0~108.7%) and reasonable intra-day and inter-day reproducibility (RSD < 9%) were obtained. This work shows that the sensitivity of colorless analyte detection-based colorimetric reaction can be effectively enhanced by integration of ES on a PAD. Schematic of the experimental setups (left) and the corresponding images (right) of the actual portable device.