已核实 DOI PubMed 相关产品: 19

A voltammetric immunosensor for clenbuterol based on the use of a MoS2-AuPt nanocomposite

Journal 2018 · IF 5.705 · 185(4): 1-10 · 发表于 2018 年 3 月 6 日
作者: Ji Renyue,Chen Shuai,Xu Wei,Qin Zhen,Qiu Jing Fu,Li Chao Rui
阅读全文 在 PubMed 查看

摘要

An ultrasensitive immunosensor for the direct detection of the illegally used livestock feed clebuterol (CLB) is described. It is based on the use of a glassy carbon electrode modified with an MoS 2 -AuPt nanocomposite and on biotin-streptavidin interaction. The use of MoS 2 -AuPt accelerates electron transfer, and this leads to a sharp increase in the electrochemical signal for the electrochemical probe hydrogen peroxide. Differential pulse voltammetry was used to record the current signal at a peak potential of −0.18 V (vs SCE). Under optimal conditions, the electrode has a linear response in the 10 pg·mL −1 to 100 ng·mL −1 CLB concentration range and a 6.9 pg·mL −1 detection limit (based on the 3σ criterium). This immunosensor is sensitive, highly specific and acceptably reproducible, and thus represents a valuable tool for the determination of CLB in pork. Schematic of a voltammetric immunosensor for the determination of clenbuterol (CLB) based on the use of a nanocomposite prepared from molybdenum disulfide and a gold-platinum alloy (MoS 2 -AuPt), and making use of the biotin-streptavidin system.

本研究中使用的阿拉丁产品

19 款产品在本文中被引用