Determination of trace selenium in high purity tellurium by hydride generation atomic fluorescence spectrometry after solid phase extraction of a diaminobenzidine-selenium chelate

Journal 2016 · IF 3.289 · 发表于 2016 年 8 月 6 日
作者: Wang Tong,Zeng Ying,Xu Jinyong
阅读全文 在 PubMed 查看

摘要

Macroporous adsorption resin was used as the sorbent for solid phase extraction and determination of the trace Se content in high purity tellurium prior to hydride generation atomic fluorescence spectrometry analysis. Selenium was converted into an organic Se chelate using 3,3′-diaminobenzidine and was separated from the tellurium matrix by solid phase extraction . The resin was packed as a column for solid phase extraction. Under optimum conditions, trace Se can be quantitatively extracted and the tellurium matrix can be removed. The Se in the eluate was determined by hydride generation atomic fluorescence spectrometry . The limit of detection (3σ) of this method was 0.22   ng   g −   1 and the relative standard deviation (RSD, n   =   5) ranged from 2.0 to 2.5% for the three investigated tellurium samples. The proposed method was successfully applied for the determination of the trace Se content in high purity tellurium samples.