Multiwalled carbon nanotubes/gold nanocomposites-based electrochemiluminescent sensor for sensitive determination of bisphenol A

Journal 2016 · IF 3.125 · 408(25): 7173-7180 · 发表于 2016 年 7 月 29 日
作者: Guo Weiwei,Zhang Amin,Zhang Xin,Huang Chusen,Yang Dapeng,Jia Nengqin
阅读全文 在 PubMed 查看

摘要

An electrochemiluminescence (ECL) sensor for bisphenol A was proposed by using l -cysteine-functionalized multiwalled carbon nanotubes/gold nanocomposites-modified glassy carbon electrode (MWCNTs-Au/GCE) based on ECL of peroxydisulfate solution. The ECL behaviors of peroxydisulfate solution had been investigated at the chitosan/MWCNTs-Au/GCE, and bisphenol A was found to have quenching effects on the ECL of peroxydisulfate solution. Both Au nanoparticles (AuNPs) and multiwalled CNTs could promote the electron transfer and synergetically amplify the ECL signal of peroxydisulfate solution. Under the optimized conditions, the ECL signal intensity was linear with the concentration of bisphenol A in the concentration range between 0.25 and 100 μM ( R  = 0.9931) with a detection limit (S/N = 3) of 0.083 μM. The constructed ECL sensor has the advantages of simplicity, sensitivity, good selectivity, and reproducibility, exhibiting a great potential application in the determination of bisphenol A.

本研究中使用的阿拉丁产品

4 款产品在本文中被引用