已核实 DOI PubMed 相关产品: 173

Graphene-ZIF8 composite material as stationary phase for high-resolution gas chromatographic separations of aliphatic and aromatic isomers

Journal 2016 · IF 3.926 · 发表于 2016 年 7 月 9 日
作者: Xiaohong Yang,Changxia Li,Meiling Qi,Liangti Qu
阅读全文 在 PubMed 查看

摘要

This work presents the separation performance of graphene-ZIF8 (G-Z) composite material as stationary phase for capillary gas chromatography (GC). The G-Z stationary phase achieved high column efficiency of 5000 plates/m determined by n -dodecane ( k   =   1.22) at 120   °C and showed weakly polar nature. Importantly, it exhibited high selectivity and resolving capability for branched alkane isomers and aromatic positional isomers, showing clear advantages over the reported neat graphene and ZIF8. In addition, it attained high resolution for geometric cis -/ trans -isomers. The G-Z column exhibited good column thermal stability up to 300   °C and column repeatability with RSD values of retention times in the range of 0.01–0.19% for intra-day, 0.05–0.88% for inter-day and 0.66–5.6% for between-column, respectively, Moreover, the G-Z column was employed for the determination of minor impurity isomers in real reagent samples, which demonstrates its promising potential in GC applications .

本研究中使用的阿拉丁产品

173 款产品在本文中被引用