已核实 DOI PubMed 相关产品: 3

An optimized method for neurotransmitters and their metabolites analysis in mouse hypothalamus by high performance liquid chromatography–Q Exactive hybrid quadrupole-orbitrap high-resolution accurate mass spectrometry

Journal 2016 · IF 2.687 · 发表于 2016 年 1 月 14 日
作者: Zong-Lin Yang,Hui Li,Bing Wang,Shu-Ying Liu
阅读全文 在 PubMed 查看

摘要

Neurotransmitters (NTs) and their metabolites are known to play an essential role in maintaining various physiological functions in nervous system. However, there are many difficulties in the detection of NTs together with their metabolites in biological samples. A new method for NTs and their metabolites detection by high performance liquid chromatography coupled with Q Exactive hybrid quadruple-orbitrap high-resolution accurate mass spectrometry (HPLC–HRMS) was established in this paper. This method was a great development of the applying of Q Exactive MS in the quantitative analysis. This method enabled a rapid quantification of ten compounds within 18   min. Good linearity was obtained with a correlation coefficient above 0.99. The concentration range of the limit of detection (LOD) and the limit of quantitation (LOQ) level were 0.0008–0.05   nmol/mL and 0.002–25.0   nmol/mL respectively. Precisions (relative standard deviation, RSD) of this method were at 0.36–12.70%. Recovery ranges were between 81.83% and 118.04%. Concentrations of these compounds in mouse hypothalamus were detected by Q Exactive LC–MS technology with this method.

本研究中使用的阿拉丁产品

3 款产品在本文中被引用