已核实 DOI 相关产品: 5

Polyfurfural film modified glassy carbon electrode for highly sensitive nifedipine determination

Journal 2015 · IF 4.504 · 发表于 2015 年 10 月 29 日
作者: Qiang Zeng,Tianyan Wei,Min Wang,Xinjian Huang,Yishan Fang,Lishi Wang
阅读全文 在 PubMed 查看

摘要

ABSTRACT A sensitive and convenient electrochemical strategy is developed for the determination of nifedipine basing on a polyfurfural film modified glassy carbon electrode (GCE) by the one-step electropolymerization of furfural . The prepared polyfurfural film/GCE exhibits excellent electrocatalytic activity towards nifedipine. A series of experimental parameters including the pH of supporting electrolyte , accumulation time and potential for nifedipine is also considered and optimized. Under optimal conditions, the proposed nifedipine sensor has a wide linear detection range from 1   ×   10 −8 to 7   ×   10 −6   mol   dm −3 , with a low detection limit of 5   ×   10 −9   mol   dm −3 . The proposed nifedipine sensor also displays excellent selectivity, stability and reproducibility. In particular, it shows splendid analytical performance for the determination of nifedipine in real pharmaceutical and human urine samples in practice.

本研究中使用的阿拉丁产品

5 款产品在本文中被引用