已核实 DOI PubMed 相关产品: 2

Determination of autoinducer-2 in biological samples by high-performance liquid chromatography with fluorescence detection using pre-column derivatization

Journal 2014 · IF 4.258 · 发表于 2014 年 8 月 11 日
作者: Xiang-Ning Song,Hai-Bin Qiu,Xiang Xiao,Yuan-Yuan Cheng,Wen-Wei Li,Guo-Ping Sheng,Xiao-Yan Li,Han-Qing Yu
阅读全文 在 PubMed 查看

摘要

Autoinducer-2 (AI-2), as a small-molecular-weight organic molecule secreted and perceived by various bacteria, enables intra- and inter-species communications. Quantitative determination of AI-2 is essential for exploring the bacterial AI-2-related physiological and biochemical processes. However, current strategies for sensitive detection of AI-2 require sophisticated instruments and complicated procedures. In this work, on the basis of the derivatization of AI-2 with 2,3-diaminonaphthalene, a simple, sensitive and cost-effective high-performance liquid chromatography with fluorescence detector (HPLC–FLD) method is developed for the quantitative detection of AI-2. Under the optimized conditions, this method had a broad linear range of 10–14,000   ng/ml ( R 2   =   0.9999), and a low detection limit of 1.0   ng/ml. Furthermore, the effectiveness of this approach was further validated through measuring the AI-2 concentrations in the cell-free culture supernatants of both Escherichia coli and Vibrio harveyi .

本研究中使用的阿拉丁产品

2 款产品在本文中被引用