已核实 DOI PubMed 相关产品: 6

Aptamer-Based Cantilever Array Sensors for Oxytetracycline Detection

Journal 2013 · IF 5.695 · 发表于 2013 年 1 月 31 日
作者: Hui Hou,Xiaojing Bai,Chunyan Xing,Ningyu Gu,Bailin Zhang,Jilin Tang
阅读全文 在 PubMed 查看

摘要

We present a new method for specific detection of oxytetracycline (OTC) at nanomolar concentrations based on a microfabricated cantilever array. The sensing cantilevers in the array are functionalized with self-assembled monolayers (SAMs) of OTC-specific aptamer, which acts as a recognition molecule for OTC. While the reference cantilevers in the array are functionalized with 6-mercapto-1-hexanol SAMs to eliminate the influence of environmental disturbances. The cantilever sensor shows a good linear relationship between the deflection amplitude and the OTC concentration in the range of 1.0–100 nM. The detection limit of the cantilever array sensor is as low as 0.2 nM, which is comparable to some traditional methods. Other antibiotics such as doxycycline and tetracycline do not cause significant deflection of the cantilevers. It is demonstrated that the cantilever array sensors can be used as a powerful tool to detect drugs with high sensitivity and selectivity.

本研究中使用的阿拉丁产品

6 款产品在本文中被引用