已核实 DOI PubMed 相关产品: 13

Design and Performance of a Desktop Time-of-Flight Mass Spectrometer for Analyzing Metal Ions

Journal 2012 · IF 1.207 · 发表于 2012 年 8 月 1 日
作者: Changjuan Guo,Wei Gao,Zhengxu Huang,Hui Zhu,Zhong Fu,Junguo Dong,Ping Cheng,Zhen Zhou
阅读全文 在 PubMed 查看

摘要

We have described a home-made desktop orthogonal-injection time-of-flight (O-ToF) mass spectrometer combining a collisional cooling system. This O-ToF consists of a simple electrospray ion source, an atmosphere–vacuum interface, an area of transmission, including a radio-frequency only quadrupole (RF-only quadrupole, RFQ) as a collisional cooling cell and an orthogonal ToF mass analyzer. In order to detect ions of small m/z value, such as small metal ions, the RFQ has been improved to weaken the mass discrimination effect against low mass ions. Metal salt solutions were used in the experiment. The system has shown a satisfactory resolving power in the spectra (m/Am = 3500), a good mass stability, a limit of detection of 80 fg and a mass accuracy of 48 ppm. The dynamic range is found to be from 10−8 mol L−1 to 10−5 mol L−1, allowing the semi-quantitative analysis of metal ions.

本研究中使用的阿拉丁产品

13 款产品在本文中被引用