已核实 DOI PubMed 相关产品: 84

Au nanoparticles anchored carbonized ZIF-8 for enabling real-time and noninvasive glucose monitoring in sweat

Journal 2025 · IF 10.5 · 发表于 2025 年 1 月 6 日
作者: Xu Yang,Fan Wu,Haowei Huang,Guojun Zheng,Hongyu Zhang,Wenrong Cai,Junyao Li,Zheng-Zhi Yin,Yong Kong
阅读全文 在 PubMed 查看

摘要

Wearable sensors can easily enable real-time and noninvasive glucose (Glu) monitoring, providing vital information for effectively preventing various complications caused by high glucose level. Here, a wearable sensor based on nanozyme-catalyzed cascade reactions is designed for Glu monitoring in sweat. Au nanoparticles (AuNPs) are anchored to the carbonated zeolitic imidazolate framework-8 (ZIF-8-C), endowing the sensor with Glu oxidase (GOx)-like and peroxidase (POD)-like activity. A flexible screen-printed carbon electrode (SPCE) is decorated with the resultant AuNPs@ZIF-8-C, which is further modified with biocompatible and swellable calcium alginate (CA) gels for the preparation of the wearable Glu sensor. The linear range for Glu detection is 10∼300 μM with a limit of detection (LOD) of 4.99 μM, which covers the physiological Glu concentration range in human sweat (10–200 μM). The developed wearable Glu sensor can fit well with the skin tissues due to the flexibility of the SPCE, and thus it can be successfully applied in real-time and noninvasive monitoring of Glu in human sweat. Additionally, the wearable Glu sensor exhibits high antibacterial activity resulted from the generated hydroxyl radicals (·OH), enabling long-term Glu monitoring in sweat.

本研究中使用的阿拉丁产品

84 款产品在本文中被引用