已核实 DOI PubMed 相关产品: 5

Development and Validation of a New High-Performance Liquid Chromatography-Ultraviolet Assay for Quantification of Mitoxantrone in Plasma of BALB/c-nu Mice

Journal 2024 · IF 1.5 · 发表于 2024 年 3 月 15 日
作者: Tao Yanru,Zhao Hua,Xiang Yujie,Li Jin,Li Yanting,Hu Jiangling,Wang Hongmei,Jiang Xinhui
阅读全文 在 PubMed 查看

摘要

The concentration of mitoxantrone in the blood of mice was determined by a high-performance liquid chromatography-ultraviolet method with aloe-emodin as the internal standard. The separation was performed on a Hypersil BDS2 column (4.6 × 250 mm, 5 μm) as the analytical column, the mobile Phase A was acetonitrile, and B was 20-mM potassium dihydrogen phosphate (adding 1% triethylamine and adjusting the pH to 2.8 with phosphoric acid) and 4.6-mM sodium octyl sulfonate. The flow rate was 1.0 mL·min−1, the detection wavelength was 243 nm, the column temperature is 25 ± 5°C and the injection amount was 20 μL. Finally, the linear range of mitoxantrone was 5–200 μg·mL−1, and the correlation coefficient was r = 0.9999. The recovery rate of the method was 91.93–105.5%, and the extraction recovery rate was 91.45–105.5%. The intraday precision and interday precision were <3.29% (limit of detection = 0.3 μg·mL−1). The HPLC method established in this paper was simple, rapid, sensitive and accurate, and can be used to determine the content of mitoxantrone in mouse plasma after tail vein injection.

本研究中使用的阿拉丁产品

5 款产品在本文中被引用