已核实 DOI PubMed 相关产品: 18

Iridium Oxide-modified Reference Screen-printed Electrodes for Point-of-Care Portable Electrochemical Cortisol Detection

Journal 2024 · IF 5.6 · 发表于 2024 年 8 月 24 日
作者: Tong Ji,Wen Ye,Weiyu Xiao,Graham Dawson,Qiuchen Dong,Christopher Gwenin
阅读全文 在 PubMed 查看

摘要

Cortisol is a well-known stress biomarker; this study focuses on using electrochemical immuno-sensing to measure the concentration of cortisol selectively and sensitively in artificial samples. Anti-cortisol antibodies have been immobilised on polycrystalline Au electrodes via strong covalent thiol bonds, fabricating an electrochemical bio-immunosensor for cortisol detection. IrO x was then anodically electrodeposited as a reference electrode on a commercial screen-printed electrode and electrochemical impedance spectrometry (EIS) studies were used to correlate the electrochemical response to cortisol concentration and the induced changes in charge transfer resistance (R ct ). A linear relationship between the R ct and the logarithm of cortisol concentration was found in concentrations ranging from 1 ng/mL to 1 mg/mL with limit of detection at 11.85 pg/mL (32.69 pM). The modification of the reference electrode with iridium oxide has greatly improved the reproducibility of the screen-printed electrode. The sensing system can provide a reliable and sensitive detection approach for cortisol measurements.