已核实 DOI PubMed 相关产品: 2

Ratiometric fluorescence sensing NADH using AIE-dots transducers at the point of care

Journal 2024 · IF 10.7 · 发表于 2024 年 1 月 29 日
作者: Sile Deng,Xiaoju Men,Muhua Hu,Xiao Liang,Yujuan Dai,Zhengkun Zhan,Zhongchao Huang,Haobin Chen,Zhuxin Dong
阅读全文 在 PubMed 查看

摘要

Reduced nicotinamide adenine dinucleotide (NADH) has a strong impact on physiological metabolism, and its concentration is related to metabolic and neurodegenerative diseases. A more reliable and accurate detection method for NADH quantitation is needed for early disease diagnosis and point-of-care testing. Aggregation-induced emission (AIE) materials are widely used to improve the sensitivity in analytes assays due to their anti-aggregation-caused quenching property. Here we developed TPA-BQD-Py AIE-dots transducers and evaluated its performance in NADH detection. The NADH concentration-dependent ratiometric sensing was based on electron transfer from TPA-BQD-Py AIE-dots to NADH with variable fluorescence intensity at 584 nm and 470 nm, resulting in high sensitivity (limit of detection at 110 nM), photostability, selectivity, and a rapid and reversible response. We further developed the application of TPA-BQD-Py AIE-dots transducers in in vivo NADH imaging using a smartphone and digital camera, respectively, demonstrating the potential for NADH point-of-care testing.

本研究中使用的阿拉丁产品

2 款产品在本文中被引用