已核实 DOI PubMed 相关产品: 11

Self-aspiration sampling design for rapid analyses of volatile organic compounds based on atmospheric pressure chemical ionization/photoionization combined ionization source mass spectrometry

Journal 2024 · IF 2.7 · 发表于 2024 年 2 月 21 日
作者: Yuxin Wang,Hengyi Xu,Bo Sui,Hui Xi,Yingjie Fu,Wuduo Zhao,Peng Li,Shihao Sun,Dingzhong Wang,Jianxun Zhang
阅读全文 在 PubMed 查看

摘要

Development of combined mass spectrometry ionization sources has enabled expansion of the application and scope of mass spectrometry. A novel hybrid ionization system combining vacuum ultraviolet (VUV) and atmospheric pressure chemical ionization (APCI) was constructed. Gaseous samples were self-aspirated into an ionization zone through a capillary by negative pressure, generated by high-speed airflow based on the Venturi effect. Compared with APCI mode alone, the signal-to-noise ratio (S/N) in APCI/VUV mode was increased by about 276-times. To increase the ionization efficiency further, correlated experimental conditions were optimized. Four types of volatile organic compounds (VOCs) were tested to evaluate the performance of the APCI/VUV ion source. Excellent linearity and limit of detection were achieved for compounds in mixed solutions. Quantitative analyses of four VOCs (toluene, cyclohexanone, styrene and ethylbenzene) using APCI/VUV-MS were done, and the relative standard deviations (RSDs) were 1.57%, 6.30%, 4.49% and 8.21%, respectively, indicating that the APCI/VUV ionization source had excellent reproducibility. Our results demonstrated that the developed method was promising for analyzing VOCs as well as being rapid, simple, and easy to operate.

本研究中使用的阿拉丁产品

11 款产品在本文中被引用