已核实 DOI PubMed 相关产品: 215

Porous microneedle-mediated continuous glucose monitoring based on reverse iontophoresis

Journal 2025 · IF 5.6 · 发表于 2025 年 8 月 27 日
作者: Rui Yao,Olanrewaju Yaasir Olatunji,Rui Wang,Zhenyu Wu,Guohua Jiang
阅读全文 在 PubMed 查看

摘要

Measuring blood glucose concentrations via continuous glucose monitoring is an important aspect of management for patients with diabetes. In this study, a continuous glucose monitoring device has been developed by integration of porous microneedles, reverse iontophoresis and biosensor technologies. An enzyme-based glucose biosensor as working electrode was firstly fabricated by electrodepositing polydopamine (PDA) onto screen-printed electrodes, and following depositing glucose oxidase (GOx) by linking of glutaric dialdehyde. The interstitial fluid (ISF) extraction could be accelerated with assistance of reverse iontophoresis. The glucose in the ISF could react with GOx on the electrode to generate a current response of blood glucose concentration. The minimally measurement results for blood glucose showed an enhanced sensitivity of 0.934 μA/mM with a high correlation (R 2 >0.99), low limit of detection (LOD ≈ 19.3 μM), broad linear detection range (LDR = 2–18 mM) that satisfices glucose detection application.

本研究中使用的阿拉丁产品

215 款产品在本文中被引用