已核实 DOI PubMed 相关产品: 10

Anti-sulfur poisoning and highly sensitive portable detection instrument for monitoring of hydrogen sulfide on-site

Journal 2025 · IF 11.3 · 发表于 2025 年 4 月 14 日
作者: Wenjing Wang,Shenghong Li,Liumo Jiang,Xiaoju Cui,Yan Gao,Haijing Ning,Qinglian Li,Xiaolin Jiang,Mo Zhang,Dehui Deng,Yang Shu,Jianhua Wang,Yafeng Guan,Xuhui Geng
阅读全文 在 PubMed 查看

摘要

The susceptibility of sensors to sulfur poisoning is the biggest challenge when monitoring high concentrations of hydrogen sulfide (H 2 S). In order to solve this problem, a portable detection instrument based on electrochemical sensor was developed. The “chainmail” catalyst of graphene-encapsulated cobalt-nickel nanoparticles (CoNi@NGs) was synthesized and doped with conductive carbon black (CCB) as sensing material. The graphene shell could protect core metal from sulfur poisoning. The anti-sulfur poisoning performance of the developed instrument reached the best level in literature. After 60 days of intermittent detection of 100 ppm H 2 S or on-site monitoring in wastewater treatment tank, the signals only decreased by 1.3 % and 1.6 %, respectively. By doping CCB, a porous structure was constructed on the working electrode membrane (WEM), which reduced the response and recovery time by 87.9 % and 77.0 %, to 14 s and 24.5 s, respectively. In addition, the instrument had a wide detection range (20 ppb-700 ppm), high sensitivity (0.580 μA/ppm) and limit of detection (LOD) of 0.52 ppb (3σ). Finally, the instrument was applied to monitor of H 2 S in wastewater treatment tank and sewage sewer of an institute, which showed that the instrument had broad application prospects for on-site monitoring of H 2 S in real environments.

本研究中使用的阿拉丁产品

10 款产品在本文中被引用