已核实 DOI PubMed 相关产品: 0

A multistage-dialysis microdevice for extraction of cryoprotectants

Journal 2017 · IF 2.062 · 19(2): 1-14 · 发表于 2017 年 4 月 18 日
作者: Zou Lili,Li Shibo,Kang Yufeng,Liu Juanjuan,He Liqun,Sun Sijie,Gao Dayong,Qiu Bensheng,Ding Weiping
阅读全文 在 PubMed 查看

摘要

In this study, we present a multistage-dialysis microdevice (MDM) for extraction of cryoprotectants (CPAs) from a CPA-laden cell suspension. We confirmed the functions of the key designs of the MDM using a fluorescence solution, we assessed the performance of the MDM by using the MDM to unload glycerin from glycerin-loaded swine erythrocytes, and we investigated the effects of the cell suspension flow rate, glycerin concentration, cell density, and membrane pore size on the clearance efficiency of glycerin (C G ), the survival rate of cells (S C ), and the recovery rate of cells (R C ). Under the designed conditions, C G , S C , and R C reached ~60%, ~90%, and ~70%, respectively. In addition, a high flow rate causes high S C and R C but a low C G . For a low glycerin concentration, C G , S C , and R C are all high. If a low cell density or a large pore membrane is used, C G is high, whereas both S C and R C are low. This work provides insight into the development of microfluidic devices for the inline extraction of cryoprotectants from a small volume of cryopreserved cells prior to the use of the cells in lab-on-a-chip applications.