摘要
The authors describe an electrochemical immunoassay for ultrasensitive direct determination of the carcinoembryonic antigen (CEA). A nanocomposite consisting of octahedral Cu2O nanocrystals covered with gold nanoparticles was utilized to modify a glassy carbon electrode which gives a strongly enhanced chronoamperometric signal for H 2 O 2 which is used as an electrochemical probe. The morphology and elemental composition of the the nanocomposite was studied by field emission scanning electron microscopy and energy dispersive X-ray spectroscopy. In addition, staphylococcal protein A was placed on the electrode for efficient capture of antibody to further enhance the sensitivity of the assay. Under optimal conditions and at a typical working voltage of −0.4 V (vs. Ag/AgCl), the response covers the 2 pg·mL −1 to 20 ng·mL −1 CEA concentration range with a 200 fg·mL −1 lower detection limit. The method was successfully applied to the determination of CEA in (spiked) human serum. Schematic of the fabrication of an electrochemical immunosensor for ultrasensitive detection the carcinoembryonic antigen. The sensor is based on the use of a glassy carbon electrode modified with an octahedral Cu 2 O-gold nanocomposite and staphylococcal protein A for signal amplification.
危险品化学品经营许可证(带存储)