Enhanced photothermal signal detection by graphene oxide integrated long period fiber grating for on-site quantification of sodium copper chlorophyllin

Journal 2021 · IF 4.616 · 146(11): 3617-3622 · 发表于 2021 年 4 月 6 日
作者: Wenjie Li,Yinping Miao,Chengwei Fei,Hongmin Zhang,Bin Li,Kailiang Zhang
阅读全文 在 PubMed 查看

摘要

An enhanced photothermal signal detection method based on graphene oxide (GO) integrated long period fiber grating (LPFG) for on-site sodium copper chlorophyllin (SCC) quantification is proposed. SCC, as a porphyrin compound, can be photonically excited to induce a stronger photothermal effect. GO offers superior molecular adsorption and thermal conductivity properties; depositing it on the LPFG surface significantly improves the sensitivity and detection efficiency of the SCC photothermal signal, when irradiated with a 405 nm laser. The experimental results showed improved performance compared with those from uncoated LPFG, with a sensitivity of 0.0587 dB (mg L−1)−1 and a limit of detection (LOD) of 0.17 mg kg−1, which is also an order of magnitude lower than that of traditional high-performance liquid chromatography. The proposed method has potential applications in the fields of real-time food safety monitoring, environmental pollutant detection, and disease diagnosis.