已核实 DOI 相关产品: 0

Co-P decorated nanoporous copper framework for high performance flexible non-enzymatic glucose sensors

Journal 2019 · IF 3.218 · 发表于 2019 年 4 月 16 日
作者: Huang Chen,Peng Sun,Meijia Qiu,Min Jiang,Jie Zhao,Dongxue Han,Li Niu,Guofeng Cui
阅读全文 在 PubMed 查看

摘要

In response to the challenges posed by diabetes, portable glucose sensors with enzyme-free catalytic materials have received great attention in recent years. Herein, a facile strategy for fabricating inexpensive and efficient electrochemical sensors is proposed. Flexible copper clad laminate (FCCL) can be firstly decorated with nanoporous copper (NP Cu) through electrodepositing Cu-Sn alloy and dealloying the Sn element. Then Co-P is further electrodeposited onto the NP Cu/FCCL and the final Co-P/NP Cu/FCCL composite electrode exhibits high catalytic activity for glucose. The great performance of the Co-P/NP Cu/FCCL mainly derives from the enlarged surface area of the NP Cu and the high electrocatalytic activity of Co-P. This electrode exhibits a high sensitivity of 1818   μA   cm −2   mM −1 , low limit of detection of 378   nM (S/N   =   3), fast response time of 4   s, a wide linear dynamic range of 0.5   μM to 10.0   mM, good anti-interference performance, excellent repeatability and considerable stability for glucose detection. The well-designed structure for catalyzing glucose in high efficiency enable the Co-P/NP Cu/FCCL a potential candidate for future flexible electrochemical sensors.