已核实 DOI PubMed 相关产品: 0

An environmentally-benign flow-batch system for headspace single-drop microextraction and on-drop conductometric detecting ammonium

Journal 2020 · IF 5.339 · 发表于 2020 年 11 月 2 日
作者: Yongrong Jiang,Xuezhi Dong,Yuzhe Li,Yan Li,Ying Liang,Min Zhang
阅读全文 在 PubMed 查看

摘要

This work presents a lab-made automatic flow-batch system for headspace single-drop microextraction and on-drop conductometric sensing ammonium. Sample and NaOH solution are simultaneously pumped into a reaction chamber (RC), where ammonium is converted to ammonia by raising pH. The converted ammonia then diffuses into the headspace of the RC, and reacts with a 100 mM boric acid drop. The conductivity of the drop is measured by an on-drop conductivity probe, which is made by two stainless-steel contacting electrodes. The result shows that the increasing rate of conductivity has a linear relationship to the ammonium concentration in sample (R 2  = 0.9945). This method has a linear range up to 400 μM, a limit of detection 2.8 μM, a relative standard deviation of 3.0% (200 μM, n = 10) and carryover coefficient 0.028. Measurements of river waters, lake waters and wastewaters have been demonstrated. The recoveries have achieved from 99.0 to 114%. This method avoids using of harmful or odorous reagents and follows the concept of green chemistry.