已核实 DOI 相关产品: 0

Multi working mode SPR chip laboratory for high refractive index detection

Journal 2023 · IF 3.8 · 31(13): 21212-21224 · 发表于 2023 年 6 月 8 日
作者: Zhuo Ren,Chunlan Liu,Yong Wei,Chunbiao Liu,Chen Shi,Xingkai Wang,Yixiong Tang,Rui Wang,Zhihai Liu
阅读全文 在 PubMed 查看

摘要

The Fiber SPR chip laboratory has become a popular choice in biochemical detection. To meet the needs of different kinds of analytes for the detection range and number of channels of the chip, we proposed a multi-mode SPR chip laboratory based on microstructure fiber in this paper. The chip laboratory was integrated with microfluidic devices made from PDMS and detection units made of bias three-core fiber and dumbbell fiber. By injecting light into different cores of a bias three-core fiber, different detection areas of dumbbell fiber can be selected, enabling the chip laboratory to enter high refractive index detection, multi-channel detection and other working modes. In the high refractive index detection mode, the chip can detect liquid samples with a refractive index range of 1.571-1.595. In multi-channel detection mode, the chip can achieve dual parameter detection of glucose and GHK-Cu, with sensitivities of 4.16 nm/(mg/mL) and 9.729 nm/(mg/mL), respectively. Additionally, the chip can switch to temperature compensation mode. The proposed multi working mode SPR chip laboratory, based on micro structured fiber, offers a new approach for the development of portable testing equipment that can detect multiple analytes and meet multiple requirements.