已核实 DOI 相关产品: 0

Electrochemical determination of epinephrine based on Ti3C2Tx MXene-reduced graphene oxide/ITO electrode

Journal 2021 · IF 4.464 · 发表于 2021 年 5 月 29 日
作者: Zicheng Li,Yingkui Guo,Hongyan Yue,Xin Gao,Shuo Huang,Xiaohua Zhang,Yanfei Yu,Haopeng Zhang,Hengwei Zhang
阅读全文 在 PubMed 查看

摘要

EP is an important natural catecholamine neurotransmitter in the central nervous system, and the change of EP concentration will lead to many diseases. Therefore, rapid and sensitive detection of EP is of great significance for clinical research. The Ti 3 C 2 T x MXene dispersion obtained by in-situ HF etching Ti 3 AlC 2 was mixed with graphene oxide (GO) dispersion synthesized by the modified Hummers method. After freeze drying and heat-treatment, the reduced graphene oxide/Ti 3 C 2 T x MXene (GMA) was obtained and then the GMA was moved onto indium tin oxide (ITO) as the GMA/ITO working electrode for electrochemical detection of epinephrine (EP). The results are as follows: Ti 3 C 2 T x MXene with a diameter of approximately 150 nm is uniformly distributed on the surface of the 3D graphene. The three-dimensional structure of GMA facilitates the entry of bioactive molecules and increases the contact area. GMA/ITO has a high sensitivity for the determination of EP 3.74 μA·μM −1 ·cm −2 . Moreover, the electrode shows a very low detection limit, well repeatability and stability, and it can detect the EP content in real urine samples.