已核实 DOI 相关产品: 0

Graphene oxide-based composite monolith as new sorbent for the on-line solid phase extraction and high performance liquid chromatography determination of ß-sitosterol in food samples

Journal 2018 · IF 4.244 · 发表于 2018 年 4 月 22 日
作者: Beijiao Cui,Bin Guo,Huimin Wang,Doudou Zhang,Haiyan Liu,Ligai Bai,Hongyuan Yan,Dandan Han
阅读全文 在 PubMed 查看

摘要

A composite monolithic column was prepared by redox initiation method for the on-line purification and enrichment of β-sitosterol, in which graphene oxide (GO) was embedded. The obtained monolithic column was characterized by scanning electron microscopy (SEM) and nitrogen adsorption-desorption isotherm measurement, which indicated that the monolith possessed characteristics of porous structure and high permeability. Under the optimum conditions for extraction and determination, the calibration equation was y = 47.92 × −0.1391; the linear range was 0.008–1.0 mg mL −1 ; the linear regression coefficient was 0.998; the limit of detection (LOD) is 2.4 μg mL −1 ; the limit of quantitation (LOQ) was 8 μg mL −1 ; precisions for intra-day and inter-day assays presented as relative standard deviations were less than 4.3% and 6.8%, respectively. Under the selective conditions, the enrichment factor of the method was 119. The recovery was in the range of 80.40–98.00%. Moreover, the adsorption amount of the monolith was compared with silica gel-C18 adsorbent and the monolith without graphene oxide being embedded. The polymerization monolithic column showed high selectivity and good permeability, and it was successfully used as on-line solid-phase extraction (SPE) column for determination of β-sitosterol in edible oil .