摘要
This work presents a split-type photoelectrochemical (PEC) immunoassay platform for ultrasensitive determination carcinoembryonic antigen (CEA) using In 2 O 3 /In 2 S 3 nanocolumn as the photoactive matrix. Initially, In 2 O 3 /In 2 S 3 nanocolumn with high photocurrent response was obtained by vulcanizing metal-organic framework-derived hollow In 2 O 3 nanomaterials for surface modification. Subsequently, sandwich immunoreactions were conducted utilizing an alkaline phosphatase-labeled secondary antibody that served as a signal pointer. A large amount of ascorbic acid was produced by the biocatalytic reaction, which enhanced the anodic photocurrent of the In 2 O 3 /In 2 S 3 nanocolumn-modified electrode. Under optimized conditions, the In 2 O 3 /In 2 S 3 nanocolumn-based PEC immunoassay exhibited a promising photocurrent response to the target CEA in the operational scope of 0.02 ng/mL to 10 ng/mL, with a limit of detection of 8.4 pg/mL. Benefiting from the excellent photoelectrochemical response of In 2 O 3 /In 2 S 3 nanocolumn to ascorbic acid and the separated detection operation, the constructed PEC immunosensing platform has the advantages of great precision and strong anti-jamming capability, and the detection results are comparable to those of commercial ELISA kits. Collectively, this study not only offered an efficacious semiconducting surface modification strategy to enhance photocurrent response activity, but also offered a novel thought for the sensitive measurement of a marker of cancer.
危险品化学品经营许可证(带存储)