已核实 DOI 相关产品: 0

Newborn screening for 3-hydroxy-3-methylglutaric aciduria using direct analysis in real-time mass spectrometry

Journal 2018 · IF 2.112 · 54(2): 134-140 · 发表于 2018 年 11 月 23 日
作者: Zhongping Huang,Yanshuxian Liu,Xiaoli Huang,Zhenzhen Hu,Huijun Liu,Lili Wang,Charles Liu,Yiping Ren
阅读全文 在 PubMed 查看

摘要

A novel method utilizing ambient thermal desorption ionization with a direct analysis in real-time source integrated with mass spectrometry (DART-MS) was established and applied to the rapid analysis of 3-hydroxy-3-methylglutaric (3-HMG) acid in the neonatal urine. Instrument parameter settings were optimized to obtain high sensitive and accurate determination of 3-HMG acid. The use of helium gas heated to temperature of 400°C was observed to permit deprotonation, 3-HMG acid producing an abundant (M-H) − ( m/z 161) in the negative ion mode. The calibration curve was determined to be linear over the range of 0.05-5 mg/L, with the correlation coefficient r = 0.9988 and the relative standard deviations (n = 6) in the range of 1.5-11.8%. The limit of detection was 0.002 mg/L, and the limit of quantitation was 0.007 mg/L. The recoveries ranged from 88.0% to 123.1%. Four urine samples from patients and four simulated urine samples were investigated. The results of DART-MS were in agreement with the values determined using established methods at the hospitals. The proposed method demonstrated significant potential in the application of the high-throughput screening in newborn screening.