A Wearable, Dual Closed-loop Insulin Delivery System for Precision Diabetes Management

Journal 2026 · IF 26.8 · 发表于 2026 年 1 月 12 日
作者: Xuecheng He,Wei Huang,Wensheng Lin,Binbin Cui,Xinyu Tian,Jing Bai,Dingyao Liu,Ivo Pang,Hao Huang,Shixian Lin,Jixiang Zhu,Jinqiang Wang,Shiming Zhang
阅读全文 在 PubMed 查看

摘要

Effective blood glucose management is an increasing demand worldwide. Traditional solutions separate glucose detection and insulin delivery, which is less efficient compared to emerging closed-loop wearable systems controlled by continuous glucose monitors (CGMs). However, CGM-controlled systems raise new safety risks, as false CGMs readings can cause insulin overdose, which results in hypoglycemia and fatal consequences. This work proposes a concept of a dual closed-loop insulin delivery system (DuoLoop) to mitigate the risk issue of CGM-controlled systems. The first closed-loop is automated insulin delivery controlled by CGM. The second closed-loop is the controlled release of glucose-responsive insulin (GRI), whose release rate depends on actual glucose levels. A customized algorithm is trained and embedded into the wearable CGMs for edge computing. The DuoLoop system shows improved safety in preliminary in vivo test (longer normoglycemia durations, 98.82% vs 92.10%), encouraging its deployment toward precision diabetes care.

本研究中使用的阿拉丁产品

127 款产品在本文中被引用