已核实 DOI 相关产品: 23

A portable NELIBS-based paper sensor for on-site trace element analysis in sewage and soil

Journal 2026 · IF 11.8 · 发表于 2026 年 3 月 9 日
作者: Fengjing Cao,Shixiang Ma,Hongwu Tian,Yachao Liu,Zhangjin He,Jing Wu,Daming Dong
阅读全文 在 PubMed 查看

摘要

Ultrafast on-site detection of elements in polluted water and soil is crucial for farming and food safety, yet remains challenging due to complex and time-consuming pretreatment. Nanoparticle-enhanced laser-induced breakdown spectroscopy (NELIBS) enables rapid on-site detection of trace heavy metals in complex samples without pretreatment while amplifying signal. We report a paper-based sensor modified with gold nanorods embedded in metal-organic frameworks (AuNRs@MOFs) using LIBS for multiplexed trace element detection in sewage and soil. It exhibits excellent linear response and low detection limits for Cu and Cr in water (0.03 μg/L, 2.49 μg/L) and soil (0.145 mg/kg, 0.072 mg/kg), with accuracy comparable to ICP-MS in real samples. This sensor provides a powerful tool for environmental protection and food safety.

本研究中使用的阿拉丁产品

23 款产品在本文中被引用