已核实 DOI 相关产品: 0

Fabrication of extracorporeal mechanical energy harvesting and implantable triboelectric nanogenerators for self-powered medical devices

Journal 2026 · IF 12.5 · 发表于 2026 年 3 月 9 日
作者: Zhenni Liu,Hu Bian,Cong Lin,Quan Chen,Jia Liu,Zhenming Chen,Jun Chen,Dongdong Fang,Junjun Huang
阅读全文 在 PubMed 查看

摘要

Wireless acquisition of extracorporeal mechanical kinematic energy offers an attractive power source for implantable bioelectronic devices. Here, we present a self-powered power supply based on the coupled effects of triboelectrification and electrostatic induction for powering implantable medical devices. A porous bio-based chitin sheet was prepared via an acid-base etching process using crayfish shells as raw material, followed by modification with poloxamer and ethyl cellulose to fabricate a bio-based triboelectric material. We demonstrate that the sheet-shaped device can be implanted in various anatomical locations, exhibiting excellent biocompatibility. By relying on externally constructed, non-contact TENG systems, triboelectric charge is generated and output through relative motion between both in vivo and in vitro triboelectric materials. Notably, the combination of materials, implant structures, and operating frequencies significantly impacts the device output performance, offering systematic guidance for optimizing TENG device applications across diverse scenarios. When the external material is PTFE, the device, implanted 0.4 cm deep in subcutaneous tissue, generates an open-circuit voltage of 592 mV and a short-circuit current of 45.7 nC at a movement frequency of 0.7 Hz. This approach represents a significant advancement in self-powered medical devices, potentially addressing the energy limitations of implantable pacemakers and other bioelectronic devices used for therapy and sensing.