摘要
Effective blood glucose management is an increasing demand worldwide. Traditional solutions separate glucose detection and insulin delivery, which is less efficient compared to emerging closed-loop wearable systems controlled by continuous glucose monitors (CGMs). However, CGM-controlled systems raise new safety risks, as false CGMs readings can cause insulin overdose, which results in hypoglycemia and fatal consequences. This work proposes a concept of a dual closed-loop insulin delivery system (DuoLoop) to mitigate the risk issue of CGM-controlled systems. The first closed-loop is automated insulin delivery controlled by CGM. The second closed-loop is the controlled release of glucose-responsive insulin (GRI), whose release rate depends on actual glucose levels. A customized algorithm is trained and embedded into the wearable CGMs for edge computing. The DuoLoop system shows improved safety in preliminary in vivo test (longer normoglycemia durations, 98.82% vs 92.10%), encouraging its deployment toward precision diabetes care.