摘要
Respiration is a crucial physiological process in the human body, and accurate relative humidity (RH) based respiratory monitoring requires effective temperature compensation. Here, we propose a polyvinyl alcohol (PVA)–graphene oxide (GO) composite film functionalized single-mode fiber (SMF)-hollow-core Bragg fiber (HCBF)-SMF hybrid fiber sensor that integrates anti-resonance (AR) and Fabry-Pérot interferometer (FPI) mechanisms, enabling RH sensing with in-situ temperature compensation. For the AR mechanism, the proposed AR-FPI sensor achieves an RH sensitivity of 68.2 pm/%RH over 36-96 %RH and a temperature sensitivity of 18.2 pm/°C over 25-60 °C. In contrast, the FPI mechanism exhibits a negligible RH response but a measurable temperature response of 1.2 pm/°C, allowing it to serve as an internal temperature-compensation channel for RH measurement. Based on these findings, a sensitivity coefficient matrix was constructed to eliminate the crosstalk between RH and temperature. The sensor exhibits reliable RH sensing performance over a broad RH range, good linearity and stability, an estimated limit of detection (LOD) of 0.22 %RH, and response/recovery times of 14/16 s, demonstrating its feasibility for non-contact respiration monitoring through breathing-induced RH variations.
危险品化学品经营许可证(带存储)