摘要
Viscosity is an important transport-related parameter in chemical and bioanalytical systems, yet its measurement in miniaturized platforms often relies on external optical tracking or dedicated sensing structures, which can complicate integration with programmable digital microfluidics. Here, we report a calibration-based impedance sensing method for droplet viscosity estimation on an electrowetting-on-dielectric digital microfluidic (EWOD-DMF) platform. A small high-frequency excitation is applied to the top plate, while the bottom electrode is used simultaneously for droplet actuation and electrical readout. As the droplet moves across the sensing electrode, the time-varying droplet–electrode overlap produces an impedance response, and a transit-time metric is extracted from the normalized amplitude trajectory. The transit time measured by impedance sensing was validated against imaging-based overlap tracking. Using glycerol–water mixtures with viscosities ranging from 1.01 to 39.23 mPa·s at 25 °C, we obtained a linear calibration between transit time and viscosity (R² = 0.99789). The method requires neither additional on-chip sensing structures nor camera-based readout during operation, and can be readily integrated with programmable EWOD-DMF workflows. This work demonstrates a calibration-based, low-volume electrical droplet viscosity estimation strategy within the validated range of 1.01–39.23 mPa·s for Newtonian glycerol–water mixtures under the specified EWOD-DMF operating conditions. Extension to complex or non-Newtonian sample systems will require application-specific calibration and validation.